Origalys
OrigaStat (with integrated electrochemical cell) - Single-Channel Potentiostats With Impedance for Electrochemical Measurements
OrigaStat (with integrated electrochemical cell) - Single-Channel Potentiostats With Impedance for Electrochemical Measurements
Study electrochemical phenomena with a reliable potentiostat/galvanostat/impedance meter. Thanks to its integrated electrochemical cell, benefits from optimized results. A true all-in-one system, discover all its features:
Features:
- Potentiostat
- Galvanostat
- Impedance meter (10µHz - 1KHz)
- Integrated electrochemical cell: OrigaCell Kit
- Speed Controller for Rotating Electrode (RDE)
- Temperature measurement
- pH meter
- Made in France
- 5-year warranty!
The OrigaCell Kit is made up of 4 elements:
1. An orange protective bubble
- To protect your cell during transport.
- Detachable
2. An electrode holder with integrated LED
- To accommodate electrodes with short cables for optimized results
- To illuminate the cell
- Detachable
3. A beaker holder
- To accommodate the plastic or glass beaker
- Can be adjusted in height
- Detachable
4. Retention tray
- To ensure the stability of the system
- Detachable
Specifications:
OrigaStat |
|||
---|---|---|---|
|
OGS080 |
OGS100 |
OGS200 |
Potentiostat |
Yes |
||
Galvanostat |
Yes |
||
Impedancemeter |
Yes |
||
Maximum current |
±100 mA |
±2 A |
|
Compliance voltage |
±17.5 V |
±35 V |
|
Max. applied potential |
±5 V |
±15 V |
|
Potential resolution |
0.003 % (30 μV) |
0.003 % (91 μV) |
|
Potential accuracy |
< 0.1% FSR (Full Scale Range) |
||
Voltages ranges |
±1 V, ±2 V and ±5 V |
±3 V, ±6 V and ±15 V |
|
Maximum scan rate |
200 V/s |
||
Current ranges |
9 |
9 (12 with low current option) |
9 (14 with low current option) |
with standard board |
±1 nA to ±100 mA |
±20 nA to ±2 A |
|
with low current option |
Not available |
1 pA to 10 nA |
|
Current accuracy |
< 0.1% FSR (Full Scale Range) |
||
Current resolution |
0.003% FSR (best resolution: 30 fA) |
0.003% FSR (best: 600 fA) |
|
Potentiostat rise/fall time |
< 2 μs |
||
Input impedance |
> 1 TΩ (//20 pF) |
||
Interfaces |
USB 2.0 |
||
Bandwidth |
1 MHz |
||
Acquisition time |
> 100 μs |
||
IR compensation |
« Manual » and « automatic feedback » |
||
Electrodes connections |
2, 3, 4 |
||
A/D converter |
16 bits |
||
EIS capability |
10 μHz to 1 KHz. |
10 μHz to 1 KHz. Up to 5 MHz with OGFEIS |
|
Analog I/O |
Not available |
Yes, 1 |
|
External current booster |
Not available |
From 5 A to 20 A |
|
Floating option |
Versatile connectivity |
||
Filters |
1μs to 1s, analog |
||
Dimensions (DxWxH) Dimensions (Unfolded feet) |
326 x 135 x 418 mm 326 x 247 x 418 mm |
400 x 135 x 418 mm 400 x 247 x 418 mm |
|
Power requirements |
90-264 Vac, 47-63 Hz, 30 VA |
90-264 Vac, 47-63 Hz, 120 VA |
|
Weight |
5.5 kg |
8 kg |
|
PC software |
OrigaMaster (USB 2.0) |
||
Cell cable length |
On demand |
||
Temperature control |
Not available |
-10°C to 105°C (14°F to 221°F) |